FDAFDA · Food and drugs·Recall number Z-0883-2013

Philips Medical Systems, GEMINI TF Big Bore, Model #882476, Positron Emission Tomography (PET) and X-Ray Computed Tomography (CT) scanning system, a diagnostic imaging systems for fixed or mobile installations that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems.

Published March 6, 2013 · Recall initiated: October 24, 2012

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software anomaly. Philips determined that the TumorLOC software program for the referenced PET/CT X-Ray scanning systems has the capability to generate Intensity Projection data sets from respiratory gated data. A circumstance was identified by the firm whereby these generated datasets are 'flipped' and incorrectly labeled left-to-right when the are saved to disk. If the flipped Intensity Project

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Medical Systems (Cleveland) Inc
Distributor
Worldwide Distribution - USA including AK, AK, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, OA MA, MD MI, MN, MO, MS, NC, D, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV; Internationally to Canada and other countries..
Lot
The affected GEMINI Big Bore system is identified with the following Serial Numbers: 9004, 9005, 9006, 9008 ,9009, 9010, 9011, 9012, 9013, 9014, 9014, 9016, 9017, 9018, 9019, 9020, 9021, 9022, 9023, 9201, 9202, 9203, 9204, 9205, 9205, 9206, 9207, 9208, 9209, 9210, 9213, 9214, 9215, 9216, 9217, 9218, 9219, 9220, and 9221.
Risk classification
Class II
Area (FDA)
device
Country
United States