SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x
Published April 13, 2022 · Recall initiated: January 19, 2022
Summary
FDA records medical device Class II recall Z-0885-2022, published April 13, 2022 and initiated January 19, 2022. Recalling firm: Volcano Corporation; Status: Ongoing; Reported quantity: 3,216; Lot: All serial numbers, limited to cases when: The SyncVision runs the current supported software version 4.2.x, AND The secondary modality (IntraSight IVUS) runs software version 5.x. Device Identifier (DI): 00845225012434. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Volcano Corporation
- Distributor
- US: WI, CA, CO, MO, ID, DC, LA, VA, KS, OK, GA, KY, IL, AZ, NY, IN, FL, TX, NC, OH, WA, OR, NM, MT, IA, MI, TN, MN, PA, AL, MA, WV, NV, NH, CT, NE, NJ, HI, AR, UT, MD, SC, MS, AK, DE OUS: JP, SA, GB, DE, PT, SG, PL, ZA, NL, MT, ES, IN, GR, EG, FI, HK, FR, NO, DK, BE, AT, IT, CO, KW, MX, PS, QA, CY, CA, TW, LV, SE, HR, SI, MY, KR, IL, TH, AU, BR, CL, SK, CH, LB, OM, ID, PK, IE, NZ, PR, RO, MM, AR, RS, RU, PA, KE, UZ
- Lot
- All serial numbers, limited to cases when: The SyncVision runs the current supported software version 4.2.x, AND The secondary modality (IntraSight IVUS) runs software version 5.x. Device Identifier (DI): 00845225012434
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAlinity m HBV AMP Kit (US and CE)April 13, 2022
- FDAUltima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.April 13, 2022
- FDAVolcano Imaging System s5i REF 807400001; CORE Mobile Imaging System (120V) REF 400-0100.01; CORE MobileDecember 6, 2017
- FDAVolcano s5, s5i, CORE and CORE Mobile systems with software version 3.4 and v3.4 software kits. Cardiology:June 1, 2016
- FDAVolcano R100 Sterile Equipment Cover; Sterile; General Hospital: These Sterile Equipment covers are providedMarch 9, 2016
- FDAVolcano s5/s5i/CORE Intravascular Ultrasound Systems software versions 3.3 and 3.4 (with and without iFR ScoutJuly 29, 2015
Questions about this recall
Was SyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.x recalled?
Yes. FDA published recall Z-0885-2022 on April 13, 2022. Reported hazard: If FFR measurement(s) are made prior to an iFR/FFR co-registration/pullback in the same procedural session, this could result in incorrect iFR/FFR Co-Registration results to be displayed, that may cause users to mistakenly use incorrect measurements, leading to inappropriate patient treatment. Image acquisition and processing system operators manual to be updated to include workflow alternatives.
What should I do if I have this product?
Voluntary: Firm initiated · Letter