Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.
Published April 13, 2022 · Recall initiated: February 9, 2022
Summary
FDA records medical device Class II recall Z-0887-2022, published April 13, 2022 and initiated February 9, 2022. Recalling firm: Maquet Cardiovascular, LLC; Status: Ongoing; Reported quantity: 60; Lot: UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Maquet Cardiovascular, LLC
- Distributor
- Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.
- Lot
- UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe Vasoview Hemopro 2 Endoscopic Vessel Harvesting System (Model VH-4000; UDI 00607567700406) is a sterile,July 22, 2026
- FDAVasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 EndoscopicOctober 1, 2025
- FDAAcrobat-i Positioner. Model Number C-XP-5000ZNovember 27, 2024
- FDAAcrobat SUV Vacuum Stabilizer System, ST. Model Number C-OM-9000ZNovember 27, 2024
- FDASyncVision Systems, Model: 400-0100.10, PN: 30000485688x, with software version 4.2.xApril 13, 2022
- FDABrius Patient Specific Brackets (components in a set of custom metal orthodontic devices); BRIUS , RX Only,April 13, 2022
Questions about this recall
Was Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001. recalled?
Yes. FDA published recall Z-0887-2022 on April 13, 2022. Reported hazard: Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.
What should I do if I have this product?
Voluntary: Firm initiated · Letter