FDAFDA · Food and drugs·Recall number Z-0887-2022
Ultima Activator II Reusable Drive Mechanism, part # C-UA-5001. Labeled REF UA-5001.
Published April 13, 2022 · Recall initiated: February 9, 2022
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Potential for corrosion on the pins of the Ultima Activator II Reusable Drive Mechanism. Harms may include allergic reaction, metal toxicity, and other delayed responses.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Maquet Cardiovascular, LLC
- Distributor
- Worldwide - United States (AL, AZ, CA, FL, IL, KY, LA, MO, MT, NM, OH, PA, TX, WI), Germany, Republic of Korea (South Korea), Russia, Thailand, United Arab Emirates.
- Lot
- UDI# 00607567700901, Lots (3): 25153700, 25157489, 25158616. Please note affected lot numbers (25153700, 25157489, 25158616) appear on the device package only. If the packaging has been discarded, the affected devices can be identified by Device Component Lot number (200501), which is etched onto all 3 device components on side of handle and on underside above each pin.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States