Mammomat Inspiration with Tomosynthesis functionality

Published March 2, 2016 · Recall initiated: February 12, 2016

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

A potential system fatal error may occur during tomosynthesis reconstruction with large breasts that cover nearly the entire detector surface and with a breast thickness larger than 90 mm. The data to be processed for the tomosynthesis application can exceed the capacity of the graphical processing unit (GPU). If the fatal error message occurs, reconstruction of the data will be aborted. The syste

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Nationwide Distribution including Puerto Rico and to the states of : AR, CA, CO, FL, GA, IA, IL, IN, KS, MA, MD, MI, MO, NC, NE, NH, NJ, NY, OH, PA, SC, TN, TX, VA and WI.
Lot
Model number: 10140000
Risk classification
Class II