Mammomat Inspiration with Tomosynthesis functionality
Published March 2, 2016 · Recall initiated: February 12, 2016
Summary
FDA records medical device Class II recall Z-0888-2016, published March 2, 2016 and initiated February 12, 2016. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 76 units; Lot: Model number: 10140000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
A potential system fatal error may occur during tomosynthesis reconstruction with large breasts that cover nearly the entire detector surface and with a breast thickness larger than 90 mm. The data to be processed for the tomosynthesis application can exceed the capacity of the graphical processing unit (GPU). If the fatal error message occurs, reconstruction of the data will be aborted. The syste
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Siemens Medical Solutions USA, Inc
- Distributor
- Nationwide Distribution including Puerto Rico and to the states of : AR, CA, CO, FL, GA, IA, IL, IN, KS, MA, MD, MI, MO, NC, NE, NH, NJ, NY, OH, PA, SC, TN, TX, VA and WI.
- Lot
- Model number: 10140000
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAARTIS icono floor. Model Number: 11327700.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono ceiling. Model Number: 11328100.July 8, 2026
- FDAARTIS icono biplane. Model Number: 11327600.July 8, 2026
- FDAStryker, T2 Fixation K-Wire 3 x 285mm WCH coated Product Usage: Intended for the temporary stabilization ofMarch 2, 2016
- FDAOlympus, Model No. GIF-H180, flexible scope, large diameter, video gastroscopeMarch 2, 2016
Questions about this recall
Was Mammomat Inspiration with Tomosynthesis functionality recalled?
Yes. FDA published recall Z-0888-2016 on March 2, 2016. Reported hazard: A potential system fatal error may occur during tomosynthesis reconstruction with large breasts that cover nearly the entire detector surface and with a breast thickness larger than 90 mm. The data to be processed for the tomosynthesis application can exceed the capacity of the graphical processing unit (GPU). If the fatal error message occurs, reconstruction of the data will be aborted. The syste
What should I do if I have this product?
Voluntary: Firm initiated · Letter