FDAFDA · Food and drugs·Recall number Z-0900-2013
Philips Digital Diagnost Stationary radiographic system
Published March 13, 2013 · Recall initiated: October 30, 2012
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
When the operator for a wall stand view selects an "image rotation" different from default, or such image rotation is preset in the examination database, then the resulting image may be rotated in the wrong direction. When the operator manually rotates the image back, electronic side markers (if programmed) are rotated also and as a result may be placed in a wrong position inside the image.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) and Internationally to Argentina, Australia, Austria, Canada, Chile, Czech republic, Estonia,France, Germany, Hungary, Inida, Ireland, Italy, Japan, Libyan Arab Jamahari, Mexico, Netherlands, New Zealand, Norway , Oman, Panama, Peru, Poland, Qatar, Reunion, Russia, Saudi Arabia, South Africa, South Korea,Spain, Sweden, Swizerland, Thailand, UAE, and the UK.
- Lot
- All Digital Diagnost systems with Eleva software version 3.0.x
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States