BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A
Published February 14, 2024 · Recall initiated: December 15, 2023
Summary
FDA records medical device Class I recall Z-0929-2024, published February 14, 2024 and initiated December 15, 2023. Recalling firm: Philips North America; Status: Ongoing; Reported quantity: 302 US; 264 OUS; Lot: All serial numbers are affected. No UDI-DI available for this product. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips North America
- Distributor
- Domestic distribution nationwide. International distribution worldwide.
- Lot
- All serial numbers are affected. No UDI-DI available for this product.
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAModel Number: 712010.November 4, 2026
- FDAPhilips Smart-hopping 2.0 AP 1.4 GHz; Product/Model Number: 867216;October 28, 2026
- FDAIngenia Ambition S and Ingenia Ambition X.; Model number (REF): 1. 781359 (S), 2. 782108 (S), 3. 782139 (S),October 7, 2026
- FDASmartPath to MR 7700; Model number (REF): 1. 782178, 2. 782207; Software Versions: R11, R12 and R13;October 7, 2026
- FDASenographe PristinaFebruary 14, 2024
- FDABrightView X, gamma camera for SPECT; Model Nos.: 6-digit format 882478; 12-digit format 453560824741February 14, 2024
Questions about this recall
Was BrightView, gamma camera system; Model Nos.: 6-digit format 882480; 12-digit format 453560279781 453560279791 453560279811 453560279801; 4x4 digit format 2170-3000A 2170-3001A 2170-3002A 2170-3003A recalled?
Yes. FDA published recall Z-0929-2024 on February 14, 2024. Reported hazard: Detector may unexpectedly fall due to a component failure. If the detector is positioned below center of gantry, there is a potential for abrasion, contusion, laceration, and/or fracture to the patient's lower limbs, as well as an interruption to normal system operation. If detector is positioned above center of gantry, there may be an interruption to normal system operation.
What should I do if I have this product?
Voluntary: Firm initiated · Letter