TotalCare SpO2RT2 bed. Product Usage: Bed, flotation therapy, powered bed may be used in a variety of settings including, but not limited to, acute care, including critical care, step down/progressive care, medical/surgical, high acuity sub-acute care, post anesthesia care unit (PACU), and sections of the emergency department (ED). The TotalCare SpO2RT¿ 2 Bed System is capable of being used with a broad patient population as determined appropriate by the caregiver or institution.

Published March 20, 2013 · Recall initiated: February 26, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

In certain situations a software problem with the TotalCare SpO2RT¿ 2 bed may occur that can lead to inadvertent loss of sufficient surface cushion pressure to adequately support the patient. This may increase the likelihood of skin irritation for patients at higher risk of developing pressure-related skin conditions. A correction to the software program is currently being developed and tested,

What to do

Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Recall details

Manufacturer
Hill-Rom, Inc.
Distributor
USA Nationwide Distribution including the state of Canada
Lot
S/N between N230AM1358 & O023AM3687.
Risk classification
Class II