The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The instrument can perform screening for abnormal samples with a high degree of accuracy, so it increases automation and efficiency in laboratories.
Published February 19, 2014 · Recall initiated: March 31, 2013
Summary
FDA records medical device Class II recall Z-0949-2014, published February 19, 2014 and initiated March 31, 2013. Recalling firm: Sysmex America, Inc.; Status: Terminated; Reported quantity: 214 units; Lot: Software versions 1.0, 1.01, 1.1.1, 1.1.2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system p
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Sysmex America, Inc.
- Distributor
- Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IS, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and WI and the country of Canada.
- Lot
- Software versions 1.0, 1.01, 1.1.1, 1.1.2
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAThe CF-70 instrument (product code: LXG; Regulation Number: 862.2050) receives magazines containing stainedApril 22, 2020
- FDASysmex CV-11 Sample Unit, an integrated modular sampler that conveys specimens to either the UF-5000 or UD-10April 22, 2020
- FDAVital CARE(TM) REPS INC. ULTRASOUND PROBE COVER, BUR HOLE COVER, 5 x 96, REF VPC0699, STERILE EO, QTY 20,February 19, 2014
- FDAStryker Orthopaedics Navigation Compatible Accolade Broach Handle 1 single unit to a package HowmedicaFebruary 19, 2014
- FDASysmex XE-2100C Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,November 14, 2012
- FDASysmex XE-5000 Automated Hematology System; an automated differential cell counter; Sysmex Corporation, Kobe,November 14, 2012
Questions about this recall
Was The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The instrument can perform screening for abnormal samples with a high degree of accuracy, so it increases automation and efficiency in laboratories. recalled?
Yes. FDA published recall Z-0949-2014 on February 19, 2014. Reported hazard: The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system p
What should I do if I have this product?
Voluntary: Firm initiated · Letter