Back to search
FDAFDA · Food and drugs·Recall number Z-0955-2019

AF w/AS 550 x 1-14ML/HR 2ML-30LO CPNB, Product code MVBT30XL CPNB, Arrow AutoFuser Disposable Pain Control Pump

Published March 6, 2019 · Recall initiated: October 2, 2018

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Possible cracks on the body of the fill port connector, which may result in a leak. If the leak is identified while the user is filling the pump, there may be a delay while an alternative pump is located. If the leak goes undetected it may result in a reduced level of drug delivery to the patient and consequently a lower level of pain relief than that prescribed or expected. This may result in the need for other, additional analgesic drug administration.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Arrow International Inc
Distributor
Nationwide distribution.
Lot
74H1701904
Risk classification
Class II
Area (FDA)
device
Country
United States