Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion.
Published March 27, 2013 · Recall initiated: January 11, 2013
Summary
FDA records medical device Class II recall Z-0956-2013, published March 27, 2013 and initiated January 11, 2013. Recalling firm: Synthes USA HQ, Inc.; Status: Terminated; Reported quantity: 4; Lot: Lot number 7965181. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Synthes USA HQ, Inc.
- Distributor
- Distributed in the states of IN, OH, PA, and TX.
- Lot
- Lot number 7965181
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASynthes Hand Switch - Electric Pen Drive, Part Number 05.001.012 The Synthes Electric Pen Drive is indicatedJanuary 8, 2014
- FDASynthes TomoFix Plates The Synthes J5606-C TomoFix Osteotomy System is used for open and closed wedgeDecember 18, 2013
- FDASynthes Trochanteric Fixation Nail As part of the Synthes Trochanteric Fixation Nail (TFN) System, the TFNDecember 18, 2013
- FDASynthes Soft Tissue Retractor Small Extendible. Offset blade facilitates easy preparation of the epipereostealDecember 11, 2013
- FDAVascular Solutions, Inc., Guardian II and Guardian II NC Hemostasis Valve, used in cardiopulmonary bypass.March 27, 2013
- FDASiemens Dimension(R) Tacrolimus Flex(R) reagent cartridge (DF107). The TACR method is used as an in vitroMarch 27, 2013
Questions about this recall
Was Variable Angle Locking Compression Plate/VA-LCP Dorsal Distal Radius Plate, Part Number 02.115.431. Indicated for: Dorsally displaced fractures, Open joint reconstruction, and Corrective osteotomies after distal radius malunion. recalled?
Yes. FDA published recall Z-0956-2013 on March 27, 2013. Reported hazard: It was discovered that one of the VA-LCP Dorsal Distal Radius Plates was incorrectly etched as Part Number 02.115.431 (110 degree) instead of the correct Part Number 02.115.231 (90 degree).
What should I do if I have this product?
Voluntary: Firm initiated · Letter