Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution
Published February 19, 2020 · Recall initiated: November 21, 2019
Summary
FDA records medical device Class I recall Z-0958-2020, published February 19, 2020 and initiated November 21, 2019. Recalling firm: Medtronic Inc.; Status: Ongoing; Lot: All Serial Numbers ( excluding devices with black retainer rings) ***NOT DISTRIBUTED WITHIN THE US***. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Inc.
- Distributor
- US:AK,AL,AR,AZ,BM,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA,MD,ME,MI,MN,MO,MS,MT,NC,ND,NE,NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY OUS: France, Germany, UK, Italy, Netherlands, Spain, Sweden, Israel, Norway, Austria, Belgium, Denmark, Czech Republic, Finland, Switzerland, Slovakia, Ireland, Poland, S. Africa, Greece, Hungary, Saudi Arabia, United Arab Emira, New Zealand, Australia, Japan, Mexico, Colombia, Brazil, S. East Asia, Korea, Canada,
- Lot
- All Serial Numbers ( excluding devices with black retainer rings) ***NOT DISTRIBUTED WITHIN THE US***
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDA2090 Programmers with serial number prefixes PKK0 and PKK1: a) 2090, Product Description: PRGM 2090September 25, 2024
- FDACobalt XT DR MRI SureScan, Model Number DDPA2D1, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDACOBALT XT VR MRI SureScan, Model Number DVPA2D1, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDAMIRRO MRI VR SureScan, Model Number DVME3D4, Implantable Cardioverter DefibrillatorJuly 3, 2024
- FDAMedtronic MiniMed 630G System with SmartGuard, Ref/Model # MMT-1515, MMT-1715, MMT-1714, MMT-1754, andFebruary 19, 2020
- FDAHUDSON RCI COMFORT FLO Humidification System, REF 2410, humidifier nebulizer kitFebruary 19, 2020
Questions about this recall
Was Medtronic MiniMed 640G Insulin Infusion Pump, Ref#s/Model(s) - MMT-1511, MMT-1711, MMT-1512, MMT-1712, MMT-1751, and MMT-1752 *Not for US Distribution recalled?
Yes. FDA published recall Z-0958-2020 on February 19, 2020. Reported hazard: There have been reported incidents of a loose reservoir that can no longer be locked into the pump. The reservoir can become loose due to a broken or missing retainer ring that prevents a proper lock. If the reservoir is not properly locked into the pump, it could lead to over or under delivery of insulin, which could then result in hypoglycemia or hyperglycemia.
What should I do if I have this product?
Voluntary: Firm initiated · Letter