BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic visualization of the heart, blood vessels or lymphatic system during or after injection of a contrast medium. It is to be used in the diagnosis of the circulatory vascular system while being combined with a positioning device and other devices.
Published March 6, 2019 · Recall initiated: January 14, 2019
Summary
FDA records medical device Class II recall Z-0959-2019, published March 6, 2019 and initiated January 14, 2019. Recalling firm: Shimadzu Medical Systems Usa Com; Status: Terminated; Reported quantity: 8 units; Lot: Serial Numbers: 40AD58C4B001, 40AD58C55001, 0261Q80503, 0261Q80804, 0161Q84501, 0161Q83102, 3M7A16A1A001, 0161Q82302. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Shimadzu Medical Systems Usa Com
- Distributor
- US Nationwide Distribution in the states of IN, OH, IL, SC
- Lot
- Serial Numbers: 40AD58C4B001, 40AD58C55001, 0261Q80503, 0261Q80804, 0161Q84501, 0161Q83102, 3M7A16A1A001, 0161Q82302
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAEquinoxe Preserve Humeral Stem, 8mm Product Usage: The Equinoxe Preserve Stems are indicated for use inMarch 6, 2019
- FDAAlcon or Endure Ophthalmic Microscope Product Usage: Used for low magnification visualization duringMarch 6, 2019
- FDACurity All Purpose Sponges, Non-woven, 4 Ply, 4" x 4" (10.2 cm x 10.2 cm) Product Code: 8044 - Product Usage:March 6, 2019
- FDAHermes NC PTA Balloon Dilatation Catheter The balloon dilatation catheter is intended to dilate stenoses inMarch 6, 2019
- FDAAF w/AS 550 x 1-14ML/HR 5ML-60LO CPNB, Product code MVBX60XL CPNB, Arrow AutoFuser Disposable Pain ControlMarch 6, 2019
- FDASHIMADZU RADspeed Pro Catalog Number: RADspeed Pro; Stationary x-ray systemJuly 5, 2017
Questions about this recall
Was BRANSIST safire Product Usage: DIGITAL ANGIOGRAPHY SYSTEM This system is intended to be used for radiologic visualization of the heart, blood vessels or lymphatic system during or after injection of a contrast medium. It is to be used in the diagnosis of the circulatory vascular system while being combined with a positioning device and other devices. recalled?
Yes. FDA published recall Z-0959-2019 on March 6, 2019. Reported hazard: Normal operation of the device is to power up the device in the morning, register the first patient, and then perform a fluoroscopy. If, however, the user powers up the device in the morning and makes an error by starting a fluoroscopy while the first patient is still being registered, the device application will abnormally terminate and require service intervention before it can be used again. This event will not occur after the first patient procedure.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail