D-10 Dual Program, which contains Update Kit D-10 A1c program diskette. United States; Bio-Rad Laboratories, Hercules, CA 94547 The Bio-Rad D-10 Hemoglobin Testing System provides an integrated method for the separation and determination of the relative percent of specific hemoglobins (e.g., A2, F, Ale} in whole blood. The separation is based on the principles of high performance liquid chromatography.

Published February 19, 2014 · Recall initiated: November 22, 2013

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

D-10 A1c Dual Program Floppy was programmed incorrectly and cartridge injections counts are not being updated correctly when switching between the D-10 Dual Short Program and Extended Program on the D-10 Hemoglobin Testing System.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Bio-Rad Laboratories, Inc.
Distributor
Worldwide Distribution - USA (nationwide) and Internationally to Latin America, Asia Pacific, Canada, Australia, New Zealand and India.
Lot
Catalog number: 220-0201: Reorder Pack Lot No./Expiration date: 20030322 3/31/2014 20030337 3/31/2014 64000081 5/31/2014 64000331 5/31/2014 64000332 5/31/2014 64000888 08/31/2014 64001344 04/30/2014 64002239 11/30/2014 ( 64002240 08/3112014 64002241 11/30/2014
Risk classification
Class III
Area (FDA)
device
Country
United States