LifeShield Drug Library Management (DLM); LifeShield Infusion Safety Software Suite v2.2 List Number 17003-02
Published December 31, 2025 · Recall initiated: November 24, 2025
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
For certain customers, the Change Summary review option ceases to operate as expected for any migrated library after the migration has been completed. Users must use alternate methods to review changes. Failure to do so may lead in over delivery, under delivery, or delay in setup on the infusion pump.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- ICU Medical, Inc.
- Distributor
- US Nationwide distribution in the states of CA, IA, LA, NY, PA, TN, VA.
- Lot
- version 2.2.1 UDI-DI 00887709127227 Issue affects customers who migrated from LifeShield DLM v2.1x to LifeShield DLM v2.2.1 and have any version of a Plum Duo v1.1 drug library that was approved or locked for review before the migration.
- Risk classification
- Class II
- Area (FDA)
- device
- Country