BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy.
Published February 19, 2014 · Recall initiated: November 15, 2013
Summary
FDA records medical device Class II recall Z-0967-2014, published February 19, 2014 and initiated November 15, 2013. Recalling firm: Brainlab AG; Status: Terminated; Reported quantity: 25 units; Lot: Model/Catalog Code: 55830-25A Instrument Star Unit ML; Serial Numbers (US Devices): 1205112007, 1205112010, 1205112011, 1205112013, 1205112014, 1205112019, 1205112021, 1205112029, 1205112034, 1205112045; Serial Numbers (Non-US Devices): 1205112004, 1205112006, 1205112008, 1205112009, 1205112015, 1205112016, 1205112017, 1205112018, 1205112020, 1205112023, 1205112025, 1205112026, 1205112027, 1205112030, 1205112032. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Brainlab AG
- Distributor
- Worldwide Distribution - USA including CA, MN, NC, NY, OH, TX and DC. Internationally to Australia, Chile, Germany, Hungary, Italy, Panama, Qatar, Russia, Thailand, United Arab Emirates.
- Lot
- Model/Catalog Code: 55830-25A Instrument Star Unit ML; Serial Numbers (US Devices): 1205112007, 1205112010, 1205112011, 1205112013, 1205112014, 1205112019, 1205112021, 1205112029, 1205112034, 1205112045; Serial Numbers (Non-US Devices): 1205112004, 1205112006, 1205112008, 1205112009, 1205112015, 1205112016, 1205112017, 1205112018, 1205112020, 1205112023, 1205112025, 1205112026, 1205112027, 1205112030, 1205112032
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x andOctober 14, 2020
- FDASpine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 )April 24, 2019
- FDAExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point withinMarch 21, 2018
- FDAAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-rayFebruary 14, 2018
- FDANormative Data Template CD for use with the I-Portal devices. Used to measure, record, or visually display theFebruary 19, 2014
- FDAVITROS Chemistry Products Mixing Cup Arrays utilized on VITROS 250 & 350 Chemistry Systems Product Usage:February 19, 2014
Questions about this recall
Was BrainLAB's VectorVision spine is intended for use as an intraoperative image-guided localization system for minimally invasive surgery. It links a freehand probe, tracked by a passive marker sensor system to virtual computer image space on a patient's preoperative image data that is processed by a VectorVision workstation. The system is indicated for any medical condition in which the use of stereoscopic surgery may be appropriate and where a reference to a rigid anatomical structure, such as the skull, the pelvis, a long bone or vertebra can be identified relative to a CT, x-ray, or MR-based model of the anatomy. recalled?
Yes. FDA published recall Z-0967-2014 on February 19, 2014. Reported hazard: Brainlab has determined that pins of thee Instrument Star Unit ML (Calibration with ICM4) with specific serial numbers might have been made of an incorrect material that is not suitable for the necessary disinfection and sterilization of the product. If corrosion occurs and is not detected by the user, and the device is used during surgery, corroded particles of the pin could, directly or indirect
What should I do if I have this product?
Voluntary: Firm initiated · Letter