Back to search
FDAFDA · Food and drugs·Recall number Z-0970-2026

Brand Name: Affixus Antegrade Femoral Nailing System Product Name: Affixus Antegrade Femoral Nails Model/Catalog Number: 815509380 Software Version: N/A Product Description: Affixus 9 mm, Length 380 mm, Right, Trochanteric Nail Component: N/A

Published December 31, 2025 · Recall initiated: December 2, 2025

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Affected implants may have undersized distal diameter along length of the shaft, which could lead to implant fatigue fracture. Potential health risk if implant fatigue fracture occurs include adverse local tissue reaction, pain or ache, non-union or malunion, or tissue damage necessitating surgical intervention.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Zimmer, Inc.
Distributor
US Nationwide distribution in the states of AL, AZ, CA, FL, IN, KY, MA, MO, MT, NC, NM, RI, TX, UT.
Lot
Lot Code: Model No 815509380 ; UDI-DI (01)00887868143557(17)340822(10)66717554 ; Lot Number 66717554 Model No 815509380 ; UDI-DI (01)00887868143557(17)340822(10)66884564 ; Lot Number 66884564 Model No 815509380 ; UDI-DI (01)00887868143557(17)350421(10)67042197 ; Lot Number 67042197
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card