FDAFDA · Food and drugs·Recall number Z-0974-2017
OEC MiniView 6800. MDL Number: D002533 The OEC MiniView 6800 Digital Mobile C-Arm is designed to provide the physician with general fluoroscopic visualization of the patient including but not limited to surgical orthopedic and extremity imaging. The device is not intended for whole-body pediatric imaging
Published January 18, 2017 · Recall initiated: December 19, 2016
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
GE Healthcare Surgery announces a voluntary field action for the OEC MiniView 6800, OEC 9900 Elite, and OEC 9800 because the workstation power cable assembly can result in intermittent power cord failures.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- GE OEC Medical Systems, Inc
- Distributor
- Worldwide Distribution - US including U.S. States: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PUERTO RICO, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY and Internationally to Canada, Mexico, and Costa Rica
- Lot
- Manufactured between 1/1/16 to 11/1/16
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Sell products like this?