Back to search
FDAFDA · Food and drugs·Recall number Z-0974-2020

Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151

Published February 12, 2020 · Recall initiated: January 4, 2019

Summary

FDA records medical device Class II recall Z-0974-2020, published February 12, 2020 and initiated January 4, 2019. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 13388 total; Lot: 8011027 8183322 8037660 8030233 8018054 7994816. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Cook Inc.
Distributor
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Lot
8011027 8183322 8037660 8030233 8018054 7994816
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Kwart Retro-Inject Ureteral Stent Set, Global Product No. G15016 G14836 G14885 G14887 G17151 recalled?

Yes. FDA published recall Z-0974-2020 on February 12, 2020. Reported hazard: The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.

What should I do if I have this product?

Voluntary: Firm initiated · Letter