Back to search
FDAFDA · Food and drugs·Recall number Z-0980-2025

The 3.0MM Neuro Match Head Soft Touch bur (5820-107-030) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT)/Otology/Neurotology/ Otorhinolaryngology, orthopedic, and general surgical applications including maxillofacial and craniofacial surgeries. The 3.0MM Prec Neuro Match Head bur (5820-107-530) is intended to be used to cut bone and bone cement during neurosurgery, including craniotomy and spinal surgery as well as ear, nose and throat (ENT)/Otology/Neurotology/Otorhinolaryngology, orthopedic, and general surgical applications including maxillofacial and craniofacial surgeries. These burs are sold as single use devices without their own IFU.

Published January 29, 2025 · Recall initiated: June 28, 2023

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

There is potential the device inside the package may not match the product information on the outer label (e.g., label states 5820-107-030 but package contains a 5820-107-530 device).

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Stryker Corporation
Distributor
Worldwide distribution - US Nationwide and the countries of Colombia, Sweden, Netherlands, UK.
Lot
GTIN 04546540362483, Lot Number 21341057
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card