Back to search
FDAFDA · Food and drugs·Recall number Z-0997-2013

Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system

Published April 3, 2013 · Recall initiated: March 8, 2013

Summary

FDA records medical device Class II recall Z-0997-2013, published April 3, 2013 and initiated March 8, 2013. Recalling firm: Siemens Medical Solutions USA, Inc; Status: Terminated; Reported quantity: 81; Lot: Model numbers 10094135, 10094137, 10094139, 10094141 and 10280959. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Siemens Medical Solutions USA, Inc
Distributor
Nationwide Distribution including AL, CA, CO, FL, IA, IL, IN, KY, MA, MD, ME, MI, MN, MO, NC, ND, NJ, NY, OH, PA, SC, TX, UT, VA, WA, and WI and to Puerto Rico.
Lot
Model numbers 10094135, 10094137, 10094139, 10094141 and 10280959.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Artis Systems with Software Artis VC20x/VC21A/VD10x that have a DSA license installed. Angiographic x-ray system recalled?

Yes. FDA published recall Z-0997-2013 on April 3, 2013. Reported hazard: The firm has become aware of a potential issue on Artis systems with the software Artis VC20x/VC21A/VD10x that have a DSA license installed. Under certain preconditions, executing roadmap OGP (organ program) in the DSA overlay mode may lead to imprecise registration of the DSA mask image with the roadmap image displayed on the live monitor.

What should I do if I have this product?

Voluntary: Firm initiated · Letter