Medtronic Intersept Custom Tubing Pack which contain Select 3D or Select CAP Arterial Cannulae. Model 3Y90R3, the following model is only approved in Europe C300901B. Sterilized using Ethylene Oxide. Non-pyrogenic. The design and components of this product are specified by the user. This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass procedures.
Published February 26, 2014 · Recall initiated: January 30, 2014
Summary
FDA records medical device Class II recall Z-1000-2014, published February 26, 2014 and initiated January 30, 2014. Recalling firm: Medtronic Inc. Cardiac Rhythm Disease Management; Status: Terminated; Reported quantity: 19483 total (7669 US, 11814 OUS); Lot: Model 3Y90R3, Lot numbers: 12227513, 12277464, 206420618, 206618641, 206842458, 206932996, 207184816. Model C300901B, Lot number 206499164. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Medtronic Inc. Cardiac Rhythm Disease Management
- Distributor
- Worldwide distribution: US (Nationwide) including Puerto Rico, and Internationally to: Algeria, Argentina, Austria, Belgium, Bosnia And Herzegovina, Brazil, Canada, Cyprus, El Salvador, France, Georgia, Germany, Greece, Hungary, India, Ireland, Italy, Japan, Jordan, Kuwait, Netherlands, Oman, Paraguay, Poland, Portugal, Russian Federation, Saudi Arabia, Slovakia, Spain, Sweden, Switzerland, Thailand, Turkey, United Kingdom, Venezuela, and Yemen.
- Lot
- Model 3Y90R3, Lot numbers: 12227513, 12277464, 206420618, 206618641, 206842458, 206932996, 207184816. Model C300901B, Lot number 206499164.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMedtronic Temporary Pacemaker, Model 5348. Single Chamber External Pulse Generator (EPG) Temporary PacemakerJune 4, 2014
- FDAMedtronic Select 3D Arterial Cannulae. Vented, model 78624 (24 Fr.). Non-Vented Cannulae, model 78724 (24February 26, 2014
- FDAMedtronic Select 3D Arterial Cannulae. Vented, models 78618 (18 Fr.) and 78620 (20 Fr.). Non-Vented Cannulae,February 26, 2014
- FDAMedtronic Select CAP Arterial Cannulae. Non-Vented, Carmeda Coated, model CB78722 (22 Fr.). Sterilized usingFebruary 26, 2014
- FDAMagna-Fx Cannulated Screw Fixation System 7.0 mm. Item codes starting with 0011420, 11460 and 11461.February 26, 2014
- FDAMedtronic Select CAP Arterial Cannulae. Vented, models 78818 (18 Fr.), 78820 (20 Fr.), 78822 (22 Fr.).February 26, 2014
Questions about this recall
Was Medtronic Intersept Custom Tubing Pack which contain Select 3D or Select CAP Arterial Cannulae. Model 3Y90R3, the following model is only approved in Europe C300901B. Sterilized using Ethylene Oxide. Non-pyrogenic. The design and components of this product are specified by the user. This product is indicated for use in the extracorporeal circuit during cardiopulmonary bypass procedures. recalled?
Yes. FDA published recall Z-1000-2014 on February 26, 2014. Reported hazard: Medtronic is recalling all models of its Select 3D and Select CAP Arterial Cannula product families due to reports of the cannula body splitting near the suture collar.
What should I do if I have this product?
Voluntary: Firm initiated · Letter