Back to search
FDAFDA · Food and drugs·Recall number Z-1003-2013

Philips Healthcare Computed Tomography X-Ray System. These devices are whole-body computed tomography (CT) x-ray systems or sub-systems, each with a continuously rotating x-ray tube and multi-row detectors enclosed by a gantry. X-ray transmission data acquired and taken at different angles can be reconstructed into cross-sectional images. Each device also includes signal analysis and display equipment, patient and equipment supports, components, and accessories.

Published April 24, 2013 · Recall initiated: April 5, 2010

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Philips Healthcare Inc.
Distributor
Worldwide Distribution including USA (nationwide)
Lot
Gemini Dual, Gemini 16 Power, Gemini GXL, Gemini LXL, Gemini TF 16, Gemini TF Base/Ready, Gemini TF 64, Gemini TF Big Bore, Precedence SPECT/CT, Brilliance CT (6, 10,16, 16P, 40, 64, and Big Bore), Brilliance iCT, Brilliance iCT SP
Risk classification
Class II
Area (FDA)
device
Country
United States

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card