Philips Healthcare Computed Tomography X-Ray System. These devices are whole-body computed tomography (CT) x-ray systems or sub-systems, each with a continuously rotating x-ray tube and multi-row detectors enclosed by a gantry. X-ray transmission data acquired and taken at different angles can be reconstructed into cross-sectional images. Each device also includes signal analysis and display equipment, patient and equipment supports, components, and accessories.
Published April 24, 2013 · Recall initiated: April 5, 2010
Summary
FDA records medical device Class II recall Z-1003-2013, published April 24, 2013 and initiated April 5, 2010. Recalling firm: Philips Healthcare Inc.; Status: Terminated; Reported quantity: 2089 US 4036 ROW; Lot: Gemini Dual, Gemini 16 Power, Gemini GXL, Gemini LXL, Gemini TF 16, Gemini TF Base/Ready, Gemini TF 64, Gemini TF Big Bore, Precedence SPECT/CT, Brilliance CT (6, 10,16, 16P, 40, 64, and Big Bore), Brilliance iCT, Brilliance iCT SP. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Philips Healthcare Inc.
- Distributor
- Worldwide Distribution including USA (nationwide)
- Lot
- Gemini Dual, Gemini 16 Power, Gemini GXL, Gemini LXL, Gemini TF 16, Gemini TF Base/Ready, Gemini TF 64, Gemini TF Big Bore, Precedence SPECT/CT, Brilliance CT (6, 10,16, 16P, 40, 64, and Big Bore), Brilliance iCT, Brilliance iCT SP
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAMobile Diagnostic X-Ray System with Digital PanelApril 12, 2017
- FDAPhilips IntellVue Info Center iX 866023; PIIC iX Upgrade 866024; IntellVue Server iX Upgrade 866026; PIICJune 18, 2014
- FDAPhilips Easy Upgrade DR, Code No: 712086 This system is used for making X-Ray exposures for diagnostics.September 4, 2013
- FDAPhilips Healthcare MobileDiagnostwDR system. Mobile Digital radiography X-Ray System Product Usage:August 28, 2013
- FDATrader Joe's Hummus Salad Dressing 12 oz., 12 units/case, SKU 96752.April 24, 2013
- FDADeGotzen XRay unit XGenus Dental X-ray unit.April 24, 2013
Questions about this recall
Was Philips Healthcare Computed Tomography X-Ray System. These devices are whole-body computed tomography (CT) x-ray systems or sub-systems, each with a continuously rotating x-ray tube and multi-row detectors enclosed by a gantry. X-ray transmission data acquired and taken at different angles can be reconstructed into cross-sectional images. Each device also includes signal analysis and display equipment, patient and equipment supports, components, and accessories. recalled?
Yes. FDA published recall Z-1003-2013 on April 24, 2013. Reported hazard: Philips Healthcare discovered the customer was performing CTDI measurements for facility accreditation and received results that were 15 to 25% higher than Philips protocol parameters stated.
What should I do if I have this product?
Voluntary: Firm initiated · Letter