Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.
Published February 4, 2015 · Recall initiated: December 17, 2014
Summary
FDA records medical device Class II recall Z-1012-2015, published February 4, 2015 and initiated December 17, 2014. Recalling firm: Aesculap, Inc.; Status: Terminated; Reported quantity: 772; Lot: material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Aesculap, Inc.
- Distributor
- Worldwide Distribution - US Nationwide
- Lot
- material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAColumbus Revision Knee System, EnduRo Knee System Product Usage: The Columbus Revision Knee System and EnduRoMarch 9, 2016
- FDAFlexible Screw Driver SJ706R; Non Sterile; B Braun; Aesculap; Aesculap AG These instruments are intended forDecember 30, 2015
- FDAS4C Occiput Torque Wrench F/Set Screw; used to tighten the S4 Cervical Spinal System clamping screws with aAugust 12, 2015
- FDAProduct Code 2C4013, Four Lead TUR Irrigation Set, for use with UROMATIC Plastic Containers, ApproximateFebruary 4, 2015
- FDASyngo RT Therapist: The intended use of the SIEMENS branded ARTISTE, ONCOR and PRIMUS family of linearFebruary 4, 2015
- FDAAesculap Miethke Shunt System, miniNAV Valve: The Miethke Shunt System is intended to shunt cerebrospinalDecember 24, 2014
Questions about this recall
Was Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum. recalled?
Yes. FDA published recall Z-1012-2015 on February 4, 2015. Reported hazard: Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).
What should I do if I have this product?
Voluntary: Firm initiated · Letter