FDAFDA · Food and drugs·Recall number Z-1012-2015

Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum.

Published February 4, 2015 · Recall initiated: December 17, 2014

Summary

FDA records medical device Class II recall Z-1012-2015, published February 4, 2015 and initiated December 17, 2014. Recalling firm: Aesculap, Inc.; Status: Terminated; Reported quantity: 772; Lot: material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Aesculap, Inc.
Distributor
Worldwide Distribution - US Nationwide
Lot
material numbers: FV400T FV401T FV402T FV403T FV404T FV405T FV406T FV407T FV408T FV409T FV499T FV790T FV791T FV792T FV793T FV794T FV795T FV796T
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Miethke Shunt System accessories Product Usage: - The Miethke Shunt System is intended to shut cerebrospinal fluid from the lateral ventricles of the brain into the peritoneum. recalled?

Yes. FDA published recall Z-1012-2015 on February 4, 2015. Reported hazard: Some Miethke Shunt System accessories may have an incorrect MR safety symbol on the shelf box label. The label should have NO symbol but instead is marked with the MR conditional symbol. The product is MR unsafe. This information is only required to be in the Instructions for Use (IFU).

What should I do if I have this product?

Voluntary: Firm initiated · Letter