Minibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device.
Published January 18, 2017 · Recall initiated: February 19, 2013
Summary
FDA records medical device Class II recall Z-1016-2017, published January 18, 2017 and initiated February 19, 2013. Recalling firm: Hospira Inc.; Status: Terminated; Reported quantity: 48,050 units; Lot: Product List Number: 14699-28; Lot Number: 17-077-NS. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Hospira Inc.
- Distributor
- Nationwide Distribution
- Lot
- Product List Number: 14699-28; Lot Number: 17-077-NS
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAInfant 25% DEXTROSE Injection, USP, 2.5 g (250 mg/mL), 10 mL Unit of Use Single-dose Syringe per carton, RxMay 31, 2017
- FDAInfant Warmer System (IWS)January 18, 2017
- FDAOPT Croutons is a synthetic bone substitute designed for the repair of bony defects and is made of pureJanuary 18, 2017
- FDAFentanyl Citrate Injection, USP, 100 mcg Fentanyl/ 2mL, Single Dose, 50 mcg/mL, Rx only, Hospira, Inc., LakeNovember 30, 2016
- FDASymbiq Two Channel Infuser; an Rx medical device infusion pump used to administer I.V. fluids. List numberNovember 16, 2016
- FDALifeCare PCA 3, PCA Serial List Number 12384 Allows for clinician administration or self-administration ofSeptember 28, 2016
Questions about this recall
Was Minibore Extension Set, 59 Inch with Spin Lock Collar, Non-DEHP, Latex-Free Intended use: For the administration of drugs and solutions to a patient's vascular system through a vascular access device. recalled?
Yes. FDA published recall Z-1016-2017 on January 18, 2017. Reported hazard: Hospira, Inc. is voluntarily recalling one lot of list 14699-28, Minibore Extension Set, 59 inch with Spin Lock Collar, Non-DEHP, identified above due to a confirmed customer report of an occlusion of solvent in the male luer lock. This occlusion has the potential to prevent the set from being primed for use and cause a delay in therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter