Kopans Breast Lesion Localization Needle The Kopans Breast Lesion Localization Needle consists of a bevel-tip needle with centimeter markings and a hookwire. The hookwire features a burnish mark that allows the operator to ensure that the hookwire is within the needle tip during needle manipulation. The Kopans Breast Lesion Localization Needle is intended for preoperative marking of nonpalpable breast lesions.
Published March 9, 2016 · Recall initiated: January 8, 2016
Summary
FDA records medical device Class II recall Z-1028-2016, published March 9, 2016 and initiated January 8, 2016. Recalling firm: Cook Inc.; Status: Terminated; Reported quantity: 134; Lot: Catalog # DKBL-20-7.0-A Lot # 6215428 date of manufacture 09/17/2015. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Product potentially exposed to body fluids during processing. Potential adverse events that may occur as a result of the foreign matter include blood borne pathogen transmission. However, the risk of transmission of blood borne pathogens is thought to be minimal to non-existent since these devices were sterilized after the potential exposure to body fluids.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cook Inc.
- Distributor
- Nationwide distribution including AZ, CA, GA, IL, MO, MS, MT, NC, NH, NY, OH, OK, and PA..
- Lot
- Catalog # DKBL-20-7.0-A Lot # 6215428 date of manufacture 09/17/2015
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAFlexor Check-Flo Introducer - Raabe Modification, Introducer Set, RPN KCFW-5.0-38-90-RB-RAABE, GPN G13069 TheNovember 10, 2021
- FDAFlexor Check-Flo Introducer, RPN KCFW-8.0-35-55-RB-HFANL0-HC, Order Number G56233 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL0-HC, Order Number G56231 Flexor Introducers andDecember 30, 2020
- FDAFlexor Check-Flo Introducer, RPN KCFW-6.0-35-55-RB-HFANL1-HC, Order Number G56236 Flexor Introducers andDecember 30, 2020
- FDANeoCoil 3.0T GEM Flex Coil. The 3.0T GEM Flex coil system includes three sizes of coil arrays and oneMarch 9, 2016
- FDAPINPOINT Endoscopic Fluorescence Imaging SystemMarch 9, 2016
Questions about this recall
Was Kopans Breast Lesion Localization Needle The Kopans Breast Lesion Localization Needle consists of a bevel-tip needle with centimeter markings and a hookwire. The hookwire features a burnish mark that allows the operator to ensure that the hookwire is within the needle tip during needle manipulation. The Kopans Breast Lesion Localization Needle is intended for preoperative marking of nonpalpable breast lesions. recalled?
Yes. FDA published recall Z-1028-2016 on March 9, 2016. Reported hazard: Product potentially exposed to body fluids during processing. Potential adverse events that may occur as a result of the foreign matter include blood borne pathogen transmission. However, the risk of transmission of blood borne pathogens is thought to be minimal to non-existent since these devices were sterilized after the potential exposure to body fluids.
What should I do if I have this product?
Voluntary: Firm initiated · Letter