Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system.
Published February 12, 2025 · Recall initiated: January 28, 2025
Summary
FDA records medical device Class II recall Z-1043-2025, published February 12, 2025 and initiated January 28, 2025. Recalling firm: GE Medical Systems, LLC; Status: Ongoing; Reported quantity: 3149 (1697 US; 1452 OUS); Lot: GTINs: 00840682138598, 00195278840226, 00840682136709). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
What to do
FDA Mandated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- GE Medical Systems, LLC
- Distributor
- Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland, Indonesia, Iraq, Ireland, Italy, Jamaica, Japan, Jordan, Kenya, Korea (Republic of) Kuwait, Libya, Lithuania, Macedonia, Malaysia, Mexico, Morocco, Netherlands, New Zealand, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, and Vietnam.
- UPC
- 00840682138598
- Lot
- GTINs: 00840682138598, 00195278840226, 00840682136709)
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASenographe PristinaNovember 11, 2026
- FDASenographe Pristina 3DNovember 11, 2026
- FDASenographe PristinaNovember 11, 2026
- FDAOptima XR240amx, Mobile Digital Radiographic SystemJuly 22, 2026
- FDAProduct is individually packaged in opaque airtight plastic packaging with Chinese lettering on one side.February 12, 2025
- FDAregard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator.February 12, 2025
Questions about this recall
Was Definium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imaging system. recalled?
Yes. FDA published recall Z-1043-2025 on February 12, 2025. Reported hazard: The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.
What should I do if I have this product?
FDA Mandated · Letter
How do I know if my unit is part of the recall?
Compare your product details with the notice. UPC: 00840682138598.