VITEK 2 Systems and VITEK 2 with MYLA.

Published May 18, 2022 · Recall initiated: March 29, 2022

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Software issue where results sent to the LIS via HL7 format for antibiotic screen tests and synergy tests do not include the user-corrected or AES-corrected interpretation. This can potentially lead to incorrect final screen/synergy test results at the LIS.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
bioMerieux, Inc.
Distributor
U.S. distribution of the FSCA to NC. OUS distribution of the FSCA to subsidiaries in Argentina, Australia, Austria, Belgium, Brazil, Canada, Chile, China, Colombia, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Italy, Japan, Mexico, Netherlands, Norway, Philippines, Poland, Portugal, Russia, Serbia, Sweden, Singapore, South Africa, South Korea, Spain, Switzerland, Thailand, Turkey, and United Kingdom.
Lot
VITEK 2 - Versions 8.01, 9.01, 9.02 (with or without 9MR2), 9.03 (with or without #9MR2). VITEK 2 with MYLA - Versions 4.8 and 4.9.
Risk classification
Class II
Area (FDA)
device
Country
United States