Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.

Published February 1, 2017 · Recall initiated: February 19, 2016

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Villa Radiology Systems LLC
Distributor
AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
Lot
06040078 07030172 06090116 06110138 07010149 07020158 07020162 07020163 07020165 07030168 07030171 07040206 09090605 10100752 06060086 06100127 09010566 08020473 11100873 15041446
Risk classification
Class II
Area (FDA)
device
Country
United States