Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Published February 1, 2017 · Recall initiated: February 19, 2016
Summary
FDA records medical device Class II recall Z-1064-2017, published February 1, 2017 and initiated February 19, 2016. Recalling firm: Villa Radiology Systems LLC; Status: Ongoing; Reported quantity: 66 distributed in US; Lot: 06040078 07030172 06090116 06110138 07010149 07020158 07020162 07020163 07020165 07030168 07030171 07040206 09090605 10100752 06060086 06100127 09010566 08020473 11100873 15041446. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Villa Radiology Systems LLC
- Distributor
- AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
- Lot
- 06040078 07030172 06090116 06110138 07010149 07020158 07020162 07020163 07020165 07030168 07030171 07040206 09090605 10100752 06060086 06100127 09010566 08020473 11100873 15041446
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAJuno DRF; Model: 709020 Multi-function x-ray systems capable of routine radiograph and general fluoroscopyFebruary 1, 2017
- FDAApollo DRF: Model: 9784220831 Multi-function x-ray systems capable of routine radiograph and generalFebruary 1, 2017
- FDAApollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and generalFebruary 1, 2017
- FDAVelara Generator with an old Firmware version used with the Philips Integris BH5000February 1, 2017
- FDAOEC 9900 Elite. MDL Numbers: D148942, D155043February 1, 2017
- FDAOwandy Radiology I -MAX Touch 3D X-rayNovember 18, 2015
Questions about this recall
Was Apollo: Models: 9784000131, 9784000231, 9784020131, 9784020231, 9784100231, 9784120131, 9784120231, 9784130131 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures. recalled?
Yes. FDA published recall Z-1064-2017 on February 1, 2017. Reported hazard: The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
What should I do if I have this product?
Voluntary: Firm initiated · Letter