Apollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and general fluoroscopy procedures.
Published February 1, 2017 · Recall initiated: February 19, 2016
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The analysis showed that the incident was caused by a chain of four faults: 1.) The injured person overcame a physical barrier (the tabletop support arm. height ca. 60 em) to access a prohibited area not intended as a working position. 2) The directions for use (device labeling and instructions for use) did not contain an explicit exclusion of the prohibited area. 3.) The movement was activated inadvertently and continuously. and it was not released (the movement requires continuous activation). 4.) Delayed activation of the emergency stop.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Villa Radiology Systems LLC
- Distributor
- AL AZ CA CO CT DC FL IA IL IN LA MA MD MI MN MO MS NJ NY OH PA Puerto Rico TN TX UT WI
- Lot
- 13101261
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States