Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents

Published February 1, 2017 · Recall initiated: September 11, 2014

Summary

FDA records medical device Class II recall Z-1067-2017, published February 1, 2017 and initiated September 11, 2014. Recalling firm: The Binding Site Group, Ltd.; Status: Terminated; Reported quantity: N/A; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
The Binding Site Group, Ltd.
Distributor
Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.
Lot
All lots.
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Related recalls

Questions about this recall

Was Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents recalled?

Yes. FDA published recall Z-1067-2017 on February 1, 2017. Reported hazard: Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.

What should I do if I have this product?

Voluntary: Firm initiated · Letter