Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents
Published February 1, 2017 · Recall initiated: September 11, 2014
Summary
FDA records medical device Class II recall Z-1067-2017, published February 1, 2017 and initiated September 11, 2014. Recalling firm: The Binding Site Group, Ltd.; Status: Terminated; Reported quantity: N/A; Lot: All lots. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- The Binding Site Group, Ltd.
- Distributor
- Worldwide Distribution - US including n FL, OH, TN, MN, NW, MI, CA. and Internationally to Canada, France, UK, Spain, Germany, Czech Republic, Belgium, Italy, Australia, Brazil, Chile, China, Costa Rica, Kuwait, New Zealand, Oman, Pakistan, Portugal, Sweded, Taiwan, and Turkey.
- Lot
- All lots.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDABrand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20June 17, 2026
- FDAThe Optilite Clinical Chemistry Analyzer. IVD. Model Number: IE700. UDI-05051700017176 - Product Usage: TheMay 27, 2020
- FDAOPTILITE IgM Kit, REF: NK012.OPT.A, IVD, UDI: 05051700017565February 26, 2020
- FDABinding Site Human IgM Kit (for SPAPLUS), REF: NK012.S The device consists of reaction buffer, sheep anti-IgMDecember 19, 2018
- FDAApollo EZ: Model: 9784152036 Multi-function x-ray systems capable of routine radiograph and generalFebruary 1, 2017
- FDAHuman lgG4 Subclass kit for use on the Beckman Coulter AU" series This kit is intended for qualifying humanFebruary 1, 2017
Questions about this recall
Was Human Subclass Kit for use on the Beckman IMMAGE"/IMMAGE"800 Analyser Product Code LK009.IM This kit is intended for qualifying human IgG subclasses 1,2,3 and 4 immunoglobulins in serum on the Beckman Coulter AU series. Meaurement of these immunoglobulins ais in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents recalled?
Yes. FDA published recall Z-1067-2017 on February 1, 2017. Reported hazard: Customer complaints identified that samples containing extremely high IgG4 levels exhibit antigen excess and will yield spuriously low IgG4 results.
What should I do if I have this product?
Voluntary: Firm initiated · Letter