FDAFDA · Food and drugs·Recall number Z-1070-2013

IMPAX CV Reporting Cardiac Catheterization module and IMPAX CV Outbound Reporting. Facilitates digital reports for catheterization, echocardiography, and congenital heart disease.

Published April 17, 2013 · Recall initiated: March 15, 2013

Summary

FDA records medical device Class II recall Z-1070-2013, published April 17, 2013 and initiated March 15, 2013. Recalling firm: AGFA Corp.; Status: Terminated; Reported quantity: 26; Lot: Model # L9M21000. Software versions: IMPAX CV 7.8, 7.8 SU1, 7.8 SU2 and 7.8 SU3, IMPAX CV 7.8 SU2.5 with IMPAX CV Outbound Report Stylesheet 7.8_HL. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Baseline pulmonary capillary wedge (PCW) pressure values from IMPAX CV Reporting Cardiac Catheterization module were not correctly transferring into the hospital's electronic health record (EHR) system.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
AGFA Corp.
Distributor
Nationwide Distribution including the states of AK, GA, IL, KY, MA, NY, NC, RI, SC, TX, and VA.
Lot
Model # L9M21000. Software versions: IMPAX CV 7.8, 7.8 SU1, 7.8 SU2 and 7.8 SU3, IMPAX CV 7.8 SU2.5 with IMPAX CV Outbound Report Stylesheet 7.8_HL
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was IMPAX CV Reporting Cardiac Catheterization module and IMPAX CV Outbound Reporting. Facilitates digital reports for catheterization, echocardiography, and congenital heart disease. recalled?

Yes. FDA published recall Z-1070-2013 on April 17, 2013. Reported hazard: Baseline pulmonary capillary wedge (PCW) pressure values from IMPAX CV Reporting Cardiac Catheterization module were not correctly transferring into the hospital's electronic health record (EHR) system.

What should I do if I have this product?

Voluntary: Firm initiated · Letter