FDAFDA · Food and drugs·Recall number Z-1073-2013
Dade(R) Actin(R) FSL Activated PTT Reagent. For use in the determination of the activated partial thromboplastin time (APTT).
Published April 17, 2013 · Recall initiated: March 11, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Siemens has observed that the listed lot numbers show an increase in Heparin sensitivity over the shelf life which is demonstrated by testing with Dade(R) Citrol(R) Heparin Control, low and high.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Siemens Healthcare Diagnostics, Inc.
- Distributor
- Worldwide Distribution-USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV and Puerto Rico and the country of Canada.
- Lot
- Catalog numbers B4219-1 and B4219-2: lot numbers 547311, exp 3/22/2013; 547312, exp 3/22/2013 and 547316A, exp 4/28/2013.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States