Back to search
FDAFDA · Food and drugs·Recall number Z-1073-2018

PORT-A-CATH II PS/Titan, Tray PAC II. Labeled as: a. VEN LP, PU, 2.6MM, UL, 8.5FR/INTRO (1/EA); b. DL, LP, PU, 2.2MM, WING-LOCK, 7FR/INTRO (1/EA); c. DL, LP, PU, 3.2MM, WING-LOCK, 10FR/INTRO (1/EA)

Published March 21, 2018 · Recall initiated: December 4, 2017

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Certain models and lots of PORT A CATH implantable port kits may have been misassembled. The kits may incorrectly contain a 6Fr or 6.5Fr size introducer instead of the correct 8.5Fr size introducer. The Power PORT A CATH II kit may also contain the incorrect locking collars in addition to the incorrect sized introducer.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
Smiths Medical ASD Inc.
Distributor
Worldwide Distribution. US nationwide including Puerto Rico, Austria, Bahamas, Belgium, Dominican Republic, Netherlands, Spain, and Turks and Caicos.
Lot
a. Item No. 21-4071-24 (Lot No. 86X046, 87X050); b. Item No. 21-8066-24 (Lot No. 87X178); c. Item No. 21-8068-24 (Lot No. 87X179)