SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).
Published February 12, 2025 · Recall initiated: December 12, 2024
Summary
FDA records medical device Class II recall Z-1073-2025, published February 12, 2025 and initiated December 12, 2024. Recalling firm: LivaNova USA, Inc.; Status: Ongoing; Reported quantity: 48; Lot: UDI-DI: 05425025750405. All Serial Numbers less than 500,000. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- LivaNova USA, Inc.
- Distributor
- US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.
- Lot
- UDI-DI: 05425025750405. All Serial Numbers less than 500,000.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAregard Item Number: 830106006, LD01114F - Newborn Kit containing the component Neo-Tee Resuscitator.February 12, 2025
- FDAAzurion 3 M12 System Code: (1) 722063 (2) 722221February 12, 2025
- FDADefinium 656HD & Discovery XR656HD (GTINs: 00840682138598, 00195278840226, 00840682136709). X-Ray imagingFebruary 12, 2025
- FDAThe Rapid fFN Specimen Collection Tube (PN: 71550-001) is a component of the Rapid fFN Specimen Collection kitFebruary 12, 2025
- FDAProduct is individually packaged in opaque airtight plastic packaging with Chinese lettering on one side.February 12, 2025
- FDAVagus Nerve Stimulation (VNS) Therapy Leads, PerenniaDURA, Model: 303. Model 303-20 PereniaDURA Lead (2.0 mmFebruary 9, 2022
Questions about this recall
Was SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS). recalled?
Yes. FDA published recall Z-1073-2025 on February 12, 2025. Reported hazard: Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.
What should I do if I have this product?
Voluntary: Firm initiated · Letter