SenTiva, REF: 1000C, Generators Part of the VNS Therapy System with Microburst Stimulation. Used for Vagus Nerve Stimulation (VNS).

Published February 12, 2025 · Recall initiated: December 12, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Vagus nerve stimulator generators may stop delivering stimulation due to an internal, mechanically activated component that may become stuck in a closed position, which may result in patients returning to baseline seizure frequency or depressive symptoms and in the longer term, patients may be required to undergo generator replacement surgery to replace their device.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
LivaNova USA, Inc.
Distributor
US: AK, AZ, CA, CO, CT, FL, GA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NY, OH, OK, OR, PA, PN, SC, SD, TN, TX, VA, VT, WA, WI, WV, UT. OUS: AE, AR, AT, BE, BG, BR, CA, CL, CO, CR, CZ, DE, DK, GR, HR, IL, IN, IT, JO, KW, LB, MK, NL, OM, PA, PE, PL, QA, RO, RS, SA, SE, SG, SI, TR, TT, UK, ZA.
Lot
UDI-DI: 05425025750405. All Serial Numbers less than 500,000.
Risk classification
Class II
Area (FDA)
device
Country
United States