All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch® VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch® VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control.
Published April 24, 2013 · Recall initiated: March 11, 2013
Summary
FDA records medical device Class I recall Z-1074-2013, published April 24, 2013 and initiated March 11, 2013. Recalling firm: Lifescan Inc; Status: Terminated; Reported quantity: US 710,413; Lot: Part numbers: Verio IQ Systems Kits: 022 267-01, 022 267-01, 022 267-03; Verio IQ Starter Kits: 022 268 01, 022 268-02; Verio IQ Warranty Kits: 022 275 01, 022 275 02. All lots and serial numbers. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Lifescan Inc
- Distributor
- USA Nationwide Distribution
- Lot
- Part numbers: Verio IQ Systems Kits: 022 267-01, 022 267-01, 022 267-03; Verio IQ Starter Kits: 022 268 01, 022 268-02; Verio IQ Warranty Kits: 022 275 01, 022 275 02. All lots and serial numbers
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDALifeScan brand OneTouch® Select Control Solution, Part Number/Model #: 02168902. Testing of Blood GlucoseApril 24, 2013
- FDALifeScan brand OneTouch® Ultra Control Solution, Part Numbers/Model Numbers: 01045808, 02141603. Testing ofApril 24, 2013
- FDAShapeMatch Cutting Guides Product Usage: The ShapeMatch Cutting Guides are intended to be used asApril 24, 2013
- FDAAlaris PC unit model 8015 with software version 9.12 Product Usage: The Alaris PC unit is the centralApril 24, 2013
- FDADeGotzen XRay unit XGenus Dental X-ray unit.April 24, 2013
- FDAIntellispace Portal (ISP) Philips Medical Systems (Cleveland), Inc., IntelliSpace Portal (ISP) is aApril 24, 2013
Questions about this recall
Was All OneTouch Verio IQ Blood Glucose Meters sold as: Verio IQ System Kits - Meter; Verio IQ Starter kits - Meter; Verio IQ Warranty Meter. Product Usage: The OneTouch® VerioIQ Blood Glucose Monitoring System is intended to be used for the quantitative measurement of glucose (sugar) in fresh capillary whole blood samples drawn from the fingertips. The system is intended to be used by a single patient and should not be used for testing multiple patients. The OneTouch® VerioIQ Blood Glucose Monitoring System is intended for self-testing outside the body (in vitro diagnostic use) by people with diabetes at home as an aid to monitor the effectiveness of diabetes control. recalled?
Yes. FDA published recall Z-1074-2013 on April 24, 2013. Reported hazard: The Verio IQ meter will shut off and revert to set up mode at glucose values above 1023 instead of displaying EXTREME HIGH GLUCOSE.
What should I do if I have this product?
Voluntary: Firm initiated · Letter