FDAFDA · Food and drugs·Recall number Z-1077-2014

PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility.

Published March 5, 2014 · Recall initiated: December 9, 2013

Summary

FDA records medical device Class II recall Z-1077-2014, published March 5, 2014 and initiated December 9, 2013. Recalling firm: DePuy Orthopaedics, Inc.; Status: Terminated; Reported quantity: 2; Lot: Catalog Number 960015 Lot Numbers 7810268 and 7806929 (US) Lot Number 7806934 (OUS). Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.

What to do

Voluntary: Firm initiated · Telephone

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
DePuy Orthopaedics, Inc.
Distributor
Worldwide Distribution-USA including the states of IA, CA, KY, and FL, and the countries of Finland, Germany, Sweden, and Czech Republic.
Lot
Catalog Number 960015 Lot Numbers 7810268 and 7806929 (US) Lot Number 7806934 (OUS)
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was PFC SIGMA Cruciate Retaining (CR); Cemented Femoral devices; (Size 5/RT) Used during primary total knee arthroplasty to improve patient mobility. recalled?

Yes. FDA published recall Z-1077-2014 on March 5, 2014. Reported hazard: Knee implant femoral devices were found to have anomalous microstructure. Porosity and the presence of blocky carbides, within the cobalt chrome part, can decrease the overall fatigue strength. This decrease in strength could cause implants to fracture under what would normally be acceptable loads.

What should I do if I have this product?

Voluntary: Firm initiated · Telephone