RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner.

Published March 12, 2014 · Recall initiated: November 25, 2013

Summary

FDA records medical device Class II recall Z-1081-2014, published March 12, 2014 and initiated November 25, 2013. Recalling firm: Elekta, Inc.; Status: Terminated; Reported quantity: 489; Lot: R4.2, R5.0, R5.1, R6.0, R6.1. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Elekta, Inc.
Distributor
Worldwide Distribution - USA including AL, CA, CT, GA, FL, IL, IN, KS, KY, LA, MD, MA, MI, MN, MO, NJ, NY, OH, PA, TN, TX, VA, and Internationally to Argentina, Austria, Bangladesh, Belgium, Botswana, Brazil, Bulgaria, Canada, Chile, China, Colombia, Cuba, Czech Republic, Egypt, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Israel, Italy, Japan, Lebanon, Mexico, Morocco, Netherland, Nigeria, Pakistan, Panama, Peru, Philippines, Portugal, Russia, Serbia, Solvenia, South Africa, South Korea, Spain, Sweden, Switzerland, Syria, Taiwan, Thailand, Turkey, United Kingdom, Venezuela, Vietnam, and Yemen.
Lot
R4.2, R5.0, R5.1, R6.0, R6.1
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was RT Desktop The EOS RT Desktop, as with the predicate SL/SLi Series user interface, is intended to be used as the user interface for the SL/SLi Series Linear Accelerators that are used for the radiation therapy treatments of malignant neoplastic diseases, as determined by a licensed medical practitioner. recalled?

Yes. FDA published recall Z-1081-2014 on March 12, 2014. Reported hazard: During the treatment table, gantry, and collimator calibration procedures, it is possible to type incorrect values or invert a value.

What should I do if I have this product?

Voluntary: Firm initiated · Letter