Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery.
Published April 24, 2019 · Recall initiated: February 27, 2019
Summary
FDA records medical device Class I recall Z-1082-2019, published April 24, 2019 and initiated February 27, 2019. Recalling firm: Brainlab AG; Status: Terminated; Reported quantity: 148 units; Lot: Version 1.0.0.249, GTIN 04056481140328. Class I means a reasonable probability of serious adverse health consequences or death.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Brainlab AG
- Distributor
- Worldwide Distribution: US (Nationwide) and countries of: Australia, Austria, Belgium, Canada, Czech Republic, Denmark, Ecuador, France, Germany, Greece, Hong Kong, Ireland, Italy, Japan, Kazakhstan, Malaysia, New Zealand, Philippines, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand and United Kingdom.
- Lot
- Version 1.0.0.249, GTIN 04056481140328
- Risk classification
- Class I
- Area (FDA)
- device
- Country
- Germany
Related recalls
- FDAUltrasound Navigation Software 1.0 and Ultrasound Integration Software for Cranial/ENT Navigation 3.0.x andOctober 14, 2020
- FDAAddback Tablet Dietary Supplement 60 tablets per 190cc Oblong WHDPE bottle Net Wet: 85gm (60 tablet wt) UPC-April 24, 2019
- FDAicumedical ChemoLock Closed Vial Spike w/Skirt, (a) REF CL-80S-4 (4 units) (b) REF CL-80S-5 (5 units). Used inApril 24, 2019
- FDAExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point withinMarch 21, 2018
- FDAAIRO Mobile CT System, Model Number: MobiCT-32 Product Usage: The AIRO is intended to be used for X-rayFebruary 14, 2018
- FDABrainlab Navigation System Spine & Trauma 3D with component Brainlab DrapeLink Registration Kit for C-armJanuary 17, 2018
Questions about this recall
Was Spine & Trauma 3D Navigation 1.0 (subpart of the system Navigation Software Spine & Trauma 3D, Version 3.0 ) Intended Use The Spine & Trauma Navigation System is intended as an intraoperative image-guided localization system to enable minimally invasive surgery. recalled?
Yes. FDA published recall Z-1082-2019 on April 24, 2019. Reported hazard: In certain occurrences, the affected navigation software application might unexpectedly display a navigated instrument in an axial, coronal/sagittal (ACS) view representation with fixed planes in the image reconstruction (the not-updated ACS view ), instead of displaying the desirable view representation Inline View , which is commonly used for navigating invasive instruments at the spine. This could lead the surgeon to be unable to determine the position of the navigated instrument. This might occur after a crash restore or after changing between different navigation workflows during the same patient treatment.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail