Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.
Published March 5, 2014 · Recall initiated: December 18, 2013
Summary
FDA records medical device Class II recall Z-1086-2014, published March 5, 2014 and initiated December 18, 2013. Recalling firm: Stryker Spine; Status: Terminated; Reported quantity: 21 units; Lot: D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Spine
- Distributor
- US Nationwide Distribution
- Lot
- D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAAVS Anchor-C Cervical Cage System Instructions For Use (IFU)- eIFU (revision 5)-Indicated for anteriorOctober 2, 2024
- FDAStryker Radius 5.5 to 3.5 Rod to Red Connector, REF 48664530. Intended to provide additional support duringNovember 18, 2015
- FDASpecialty Audible Torque Wrench Mfg by: Stryker Spine, Catalog number IS3002XLP, Non-sterile. Used to finalDecember 24, 2014
- FDASpecialty Aria Inserter, REF IS3021ARI, Non-Sterile, This instrument is part of the Aria product brand. It isSeptember 17, 2014
- FDASiemens Artis zee Ceiling Systems. Angiographic x-ray system.March 5, 2014
- FDAVITROS Chemistry Products Na+ Slides, manufactured by Ortho Clinical Diagnostics 513 Technology BlvdMarch 5, 2014
Questions about this recall
Was Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges. recalled?
Yes. FDA published recall Z-1086-2014 on March 5, 2014. Reported hazard: Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.
What should I do if I have this product?
Voluntary: Firm initiated · Letter