CSI, Diamondback 360 Coronary Orbital Atherectomy System, Part Number 70058-02, Model Number DBEC-125 Product Usage: The system is indicated to facilitate stent delivery in patients with coronary artery disease (CAD) who are acceptable candidates for PTCA or stenting due to de novo, severely calcified coronary artery lesions.
Published March 5, 2014 · Recall initiated: January 27, 2014
Summary
FDA records medical device Class III recall Z-1095-2014, published March 5, 2014 and initiated January 27, 2014. Recalling firm: Cardiovascular Systems, Inc.; Status: Terminated; Reported quantity: 318; Lot: Lot: 85408 -85412, 85585 -85587, 87276, 87278, 87280, 87317, 87462, 87464, 87793, 87949 - 87952, 88520 - 88524, 88975 - 88979, 89498, 89501- 89504, 89952, 89954, 89956, 91876, 91877, 91891, 92124, 92156. Class III means the product is not likely to cause adverse health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Cardio vascular Systems Inc. has initiated a recall to request the immediate removal and return of Diamondback 360 Coronary Orbital Atherectomy Device (OAD). The device mistakenly contains a saline line that is used in our peripheral atherectomy device. This saline line was not part of the FDA approval for use in the coronary device. It therefore does not meet specification for this device. C
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Cardiovascular Systems, Inc.
- Distributor
- USA Nationwide Distribution in the states of AZ, CA, FL, MA, MN, MS, NC, NJ, NY, PA, TN, TX, VA.
- Lot
- Lot: 85408 -85412, 85585 -85587, 87276, 87278, 87280, 87317, 87462, 87464, 87793, 87949 - 87952, 88520 - 88524, 88975 - 88979, 89498, 89501- 89504, 89952, 89954, 89956, 91876, 91877, 91891, 92124, 92156.
- Risk classification
- Class III
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDADiamondback 360 Coronary Orbital Atherectomy System, Model Number: DBEC-125. The system is indicated toApril 1, 2015
- FDACSI Cardiovascular Systems, Inc., Diamondback 360 Peripheral Orbital Atherectomy System, Model NumberAugust 20, 2014
- FDANovaSure Impedance Controlled Endometrial Ablation System Radiofrequency Controller, Model Number: 71978-001March 5, 2014
- FDAInvivo Corporation Expression IP5 Information Portal Product Usage: The Expression Information Portal (ModelMarch 5, 2014
- FDAProduct Name: HER2 CISH pharmDx Kit Catalog/Model number: SK109, lot 20000910 Product Usage: For in vitroMarch 5, 2014
- FDACSI Stealth 360-degree Orbital Atherectomy Device (OAD), Part Number 70056-11, Model Number PRD-SC30-125S.July 18, 2012
Questions about this recall
Was CSI, Diamondback 360 Coronary Orbital Atherectomy System, Part Number 70058-02, Model Number DBEC-125 Product Usage: The system is indicated to facilitate stent delivery in patients with coronary artery disease (CAD) who are acceptable candidates for PTCA or stenting due to de novo, severely calcified coronary artery lesions. recalled?
Yes. FDA published recall Z-1095-2014 on March 5, 2014. Reported hazard: Cardio vascular Systems Inc. has initiated a recall to request the immediate removal and return of Diamondback 360 Coronary Orbital Atherectomy Device (OAD). The device mistakenly contains a saline line that is used in our peripheral atherectomy device. This saline line was not part of the FDA approval for use in the coronary device. It therefore does not meet specification for this device. C
What should I do if I have this product?
Voluntary: Firm initiated · Letter