Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to provide high flux and low flux hemodialysis, hemofiltration and ultrafiltration on patients weighing 15 Kilograms or more. The Phoenix system is to be used with either high or low permeability dialyzers. The device is intended to be used by trained operators when prescribed by a physician, in a chronic care dialysis facility or acute care unit
Published February 1, 2017 · Recall initiated: January 4, 2017
Summary
FDA records medical device Class II recall Z-1095-2017, published February 1, 2017 and initiated January 4, 2017. Recalling firm: Baxter Healthcare Corporation; Status: Terminated; Reported quantity: 256 kits; Lot: Product code: 6997274; Lot codes: 160718E, 160720E, 160825E. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Baxter Healthcare Corporation
- Distributor
- Nationwide Distribution
- Lot
- Product code: 6997274; Lot codes: 160718E, 160720E, 160825E
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDATruSystem 7500 surgical table, Model/Catalog Number: 4091000November 11, 2026
- FDATruSystem 7000 U Operating Table (dV), Product Code 1723633November 11, 2026
- FDATruSystem 7500 Hybrid Plus (FC), Fixed-Installation Operating Table Column, Model/Catalog Number: 1854087November 11, 2026
- FDAOperating Table TruSystem 7000 U, Surgical table, universal, Model/Catalog Number: 1604788November 11, 2026
- FDATryptic Soy Agar, Acumedia PN 7100 500g, 2Kg, 10Kg, and 50Kg sizesFebruary 1, 2017
- FDAHitachi Oasis MRI System - C-Spine CoilFebruary 1, 2017
Questions about this recall
Was Maintenance Kit PH2, Cod. 6997274 The Phoenix Hemodialysis delivery system is intended to be used to provide high flux and low flux hemodialysis, hemofiltration and ultrafiltration on patients weighing 15 Kilograms or more. The Phoenix system is to be used with either high or low permeability dialyzers. The device is intended to be used by trained operators when prescribed by a physician, in a chronic care dialysis facility or acute care unit recalled?
Yes. FDA published recall Z-1095-2017 on February 1, 2017. Reported hazard: Baxter Healthcare Corporation is issuing a voluntary product recall for several lots of Phoenix Dialysis Machine Preventive Maintenance Kits due to incorrect silicone straight connectors components packaged in these kits. During Preventive Maintenance, the two silicone connectors must be installed in the Phoenix mass-balance section of hydraulic circuit only if the Phoenix machine has already been modified with "CONN. IMPROVING RETR. KIT" (Code SP00721). Installation of these connectors in machines modified with SP00721 may result in fluid leaks during dialysis treatment. These Preventive Maintenance Kits include the FILTER sub-group, which contains the silicone straight connectors that are larger than expected.
What should I do if I have this product?
Voluntary: Firm initiated · Letter