Hitachi Oasis MRI System - C-Spine Coil

Published February 1, 2017 · Recall initiated: January 6, 2017

Summary

FDA records medical device Class II recall Z-1096-2017, published February 1, 2017 and initiated January 6, 2017. Recalling firm: Hitachi Medical Systems America Inc; Status: Terminated; Reported quantity: 180 units; Lot: M001 to M028, M029 to M083, M085 to M090, M092 to M104, M106, M110 to M111, M113 to M114, M116 to M123, M125, M128 to M131, M133, M139 to M141, M143 to M147, M150, M152, M154, M159, M161, M163, M166, M169 to M176, M179, M182, M183, M185, M186, M190, M194 to 196, M201, M207 to M238. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.

What to do

Voluntary: Firm initiated · Letter

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Hitachi Medical Systems America Inc
Distributor
USA (nationwide) distribution. The product was distributed to the following states: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NH, NJ, NM, NY, OH, OK, PA, PR, RI, SC, SD, TN, TX, UT, VA, WA, WI, and WY. There are three (3) government accounts for this recall: VA Medical Center/Lyster Army Health Clinic, Parma VA Medical Center, Outpatient Clinic. There are no Canada accounts for this recall. There are no Mexico accounts for this recall.
Lot
M001 to M028, M029 to M083, M085 to M090, M092 to M104, M106, M110 to M111, M113 to M114, M116 to M123, M125, M128 to M131, M133, M139 to M141, M143 to M147, M150, M152, M154, M159, M161, M163, M166, M169 to M176, M179, M182, M183, M185, M186, M190, M194 to 196, M201, M207 to M238
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Hitachi Oasis MRI System - C-Spine Coil recalled?

Yes. FDA published recall Z-1096-2017 on February 1, 2017. Reported hazard: The customer indicated that the coil balun was hot to the touch when removing the coil from the patient table after an exam.

What should I do if I have this product?

Voluntary: Firm initiated · Letter