Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems.
Published April 17, 2019 · Recall initiated: March 12, 2019
Summary
FDA records medical device Class II recall Z-1109-2019, published April 17, 2019 and initiated March 12, 2019. Recalling firm: Elekta Limited; Status: Terminated; Reported quantity: 12 units; Lot: Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Elekta Limited
- Distributor
- Worldwide Distribution - US Nationwide in the states of TX, NY, WI, and Denmark, Germany, Italy, Netherlands, Sweden, United Kingdom
- Lot
- Unity Systems: 13553-006/600003, 13371-002/600021, 10817-011/600010, 10562-011/600016, 10719-T03001/600013, 30004070-001/600023, 10420-AVL-U/600008, 11014-UMCU-U/600007, 10157-006/600014, 11611-40/600011, 30001461-J001/600002, 11489-21/600009,
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDADivalproex Sodium Extended-Release Tablets, USP, 250 mg a) 100-count bottle (NDC 55111-0533-01), b) 500-countApril 24, 2019
- FDAVolumex (Iodinated I 131 Albumin) Injection USP, 25 uCi per 1 mL syringe, Rx Only, Manufactured for DaxorApril 24, 2019
- FDAiTotal CR Total Knee Replacement System, TCR-121-1111 (Catalog No.) / TCR-121-1111-010102 (Model No.)April 17, 2019
- FDAGravity IV Set with 3-port Closed Stopcock Manifold Model CM42500E-07April 17, 2019
- FDAWalgreens Probiotic 12.5 Billion Active Cultures, 30 capsules; Dietary Supplement. Probiotic Blend 12.5April 17, 2019
- FDAImpossible Foods Bulk, 5 lb. vacuum sealed plastic-pack. Keep Frozen; This is a plant product that is looksApril 17, 2019
Questions about this recall
Was Elekta Unity, Image-Guided Radiation Therapy System Product Usage: Elekta Unity using Magnetic Resonance Imaging is indicated for radiation therapy treatments and stereotactic radiation treatments of malignant and benign diseases anywhere in the body as determined by a licensed medical practitioner in accordance with a defined treatment plan. Elekta Unity is intended for use with compatible Treatment Planning and Oncology Information Systems. recalled?
Yes. FDA published recall Z-1109-2019 on April 17, 2019. Reported hazard: Users need to be aware when using these protocols for daily on line plan adaptation that: 1) The images acquired using these protocols do not represent the average position of the anatomy during the respiratory motion cycle. The images are based on data acquired around full expiration. 2) The display of the images in the Elekta Unity Application software does not provide information about the protocol used to acquire the image eg. with or without respiratory triggering. Users must select an appropriate scan protocol that is representative of the respiratory phase used in the reference plan.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail