Product 12 consists of all product under product code JDI , and same usage: Item no: 902602935 6 DEGREE COCR FEM HEAD 28 902602835 6 DEGREE COCR FEM HEAD 28 902603235 6 DEGREE COCR FEM HEAD 32 902603335 6 DEGREE COCR FEM HEAD 32 902603600 6 DEGREE COCR FEM HEAD 36 902603610 6 DEGREE COCR FEM HEAD 36 902603614 6 DEGREE COCR FEM HEAD 36 Product Usage: For use in total or hemi hip arthroplasty
Published March 23, 2016 · Recall initiated: January 11, 2016
Summary
FDA records medical device Class II recall Z-1110-2016, published March 23, 2016 and initiated January 11, 2016. Recalling firm: Zimmer Manufacturing B.V.; Status: Terminated; Reported quantity: 209; Lot: lot no62240860 62339639 62395853 62240859 62279765 62339640 62395852 62089508 62094969 62185962 62240862 62273366 62279766 61660149 62279769 62284368 62284368N 62284371 62093453. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Zimmer Manufacturing B.V.
- Distributor
- US Nationwide Distribution
- Lot
- lot no62240860 62339639 62395853 62240859 62279765 62339640 62395852 62089508 62094969 62185962 62240862 62273366 62279766 61660149 62279769 62284368 62284368N 62284371 62093453
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAVersaFx Femoral Fixation System Hex head compression screw 4.7 mm dia. 2 inch (51 mm) length 47116202400May 25, 2016
- FDAPersona The Personalized Knee System Articular Surface Posterior Stabilized (PS) Left Height 14mm Sterile. ForApril 13, 2016
- FDAProduct 48 consists of all product under product code: JWH and same usage: Item no: 597206526 NEXGEN ALL-POLYMarch 23, 2016
- FDAProduct 25 consists of all product under product code: LPH and same usage: Item no: 234703104 DISTAL MEDIALMarch 23, 2016
- FDAProduct 10consists of all product under product code HWC, and same usage: Item no: 47484502601 4.5 X 26 CORTMarch 23, 2016
- FDAProduct 14 consists of all product under product code HSB , and same usage: Item no: 225600230 ASIA ONE-PIECEMarch 23, 2016
Questions about this recall
Was Product 12 consists of all product under product code JDI , and same usage: Item no: 902602935 6 DEGREE COCR FEM HEAD 28 902602835 6 DEGREE COCR FEM HEAD 28 902603235 6 DEGREE COCR FEM HEAD 32 902603335 6 DEGREE COCR FEM HEAD 32 902603600 6 DEGREE COCR FEM HEAD 36 902603610 6 DEGREE COCR FEM HEAD 36 902603614 6 DEGREE COCR FEM HEAD 36 Product Usage: For use in total or hemi hip arthroplasty recalled?
Yes. FDA published recall Z-1110-2016 on March 23, 2016. Reported hazard: As a result of the insufficient sealer calibration data for product packaged in firm's Building II between August 2010 and April 2013. The affected products are sterile.
What should I do if I have this product?
Voluntary: Firm initiated · Letter