FDAFDA · Food and drugs·Recall number Z-1114-2013
CoaguChek XS Pro meters (catalog number 05530199160)bar code scanner. Used with the CoagCHEK XS Pro Meters to quantitatively determines prothrombin time ("PT").
Published April 24, 2013 · Recall initiated: March 6, 2013
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Roche investigation into complaints found Accu-Chek Inform II and CoaguChek XS Pro System barcode reading meters had the potential of intermittent erroneous decoding of patient identification if barcodes being read fail to meet size and/or quality standards. Errors have been confirmed with the EAN 13 barcodes which should not be used for patient or operator identification.
What to do
Voluntary: Firm initiated · Letter
Recall details
- Manufacturer
- Roche Diagnostics Operations, Inc.
- Distributor
- Nationwide Distribution.
- Lot
- catalog number 05530199160
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States