REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants.
Published April 24, 2013 · Recall initiated: March 20, 2013
Summary
FDA records medical device Class II recall Z-1124-2013, published April 24, 2013 and initiated March 20, 2013. Recalling firm: Biomet U.K., Ltd.; Status: Terminated; Reported quantity: 430; Lot: 32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
What to do
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Biomet U.K., Ltd.
- Distributor
- Worldwide Distribution - USA (nationwide), Europe, Japan, Australia, Canada, Singapore, South Africa, and Costa Rica.
- Lot
- 32-422760 ( sold as part of instrument set 32-422763 ) ZB100901,ZB110201,ZB110301,ZB110302,ZB110303, ZB110304,ZB110601,ZB110602,ZB110701,ZB111201, ZB111202,ZB120201,ZB120202,ZB120701,ZB120801, ZB120802,ZB121102,ZB121103
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAOxford Partial Knee System Left Medial Tibial Trial Tray Size E Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size C Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Right Medial Tibial Trial Tray Size F Product Usage: Instruments for use with theMarch 1, 2017
- FDAOxford Partial Knee System Phase 3 Tibial Template Right Medial Size B Product Usage: Instruments for use withMarch 1, 2017
- FDAEVOTECH Endoscope Cleaner & Reprocessor System, P/N 50004 Designed to automatically clean and high-levelApril 24, 2013
- FDAInstrumentation Industries BE183-SU and BE 183 SU-12 Manual Jet Ventilators; BE 183-SUR and BE 183 SUR-12April 24, 2013
Questions about this recall
Was REF 32-422760 Oxford Uni Knee System,Toffee Hammer, non-sterile, found in REF 32-422763 Oxford Mlcroplasty Knee System Tibial instrument Tray small surgical hammer supplied with the Oxford Partial Knee Microplasty Instrument set, used for impacting implants. recalled?
Yes. FDA published recall Z-1124-2013 on April 24, 2013. Reported hazard: Investigation of complaints found excessive use of the instrument could result in cracking on the weld between the head and the handle, which could lead to part of the head becoming detached from the handle.
What should I do if I have this product?
Voluntary: Firm initiated · Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit