Back to search
FDAFDA · Food and drugs·Recall number Z-1141-2024

HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-08, ; SAM 500P/500-BAS-UK-10

Published February 21, 2024 · Recall initiated: February 14, 2024

Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.

What to do

Voluntary: Firm initiated · Letter

Recall details

Manufacturer
HeartSine Technologies Ltd
Distributor
Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
Lot
Type/UDI-DI or GTIN/Serial Number: SAM 350P/05060167120671, 05060167127670, +M727SAM350P0/$$+7/17D00023014, 18D00020508, 19D00017655, 20D00006524, 21D91081653, 22D91152391; SAM 360P/05060167127687, 05060167127311,5060167122927/21E91049810, 23E90001379, 23E90010409; SAM 450P/05060167129773, +M727SAM450P0/$$+7, +M727SAM450P0/$$+7/17G00001893, 18G00001553; SAM 500P/05060167122453/22B91071233
Risk classification
Class II
Area (FDA)
device
Country
United Kingdom

Sell products like this?

Recalert watches your whole catalog for recalls like this one, automatically.

Start free — no card