HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-08, ; SAM 500P/500-BAS-UK-10
Published February 21, 2024 · Recall initiated: February 14, 2024
Summary
FDA records medical device Class II recall Z-1141-2024, published February 21, 2024 and initiated February 14, 2024. Recalling firm: HeartSine Technologies Ltd; Status: Ongoing; Reported quantity: 12; Lot: Type/UDI-DI or GTIN/Serial Number: SAM 350P/05060167120671, 05060167127670, +M727SAM350P0/$$+7/17D00023014, 18D00020508, 19D00017655, 20D00006524, 21D91081653, 22D91152391; SAM 360P/05060167127687, 05060167127311,5060167122927/21E91049810, 23E90001379, 23E90010409; SAM 450P/05060167129773, +M727SAM450P0/$$+7, +M727SAM450P0/$$+7/17G00001893, 18G00001553; SAM 500P/05060167122453/22B91071233. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- HeartSine Technologies Ltd
- Distributor
- Worldwide - US Nationwide distribution in the states of TX, CA, IL, CO, FL, NC, MI and the countries of Netherlands, Japan, New Zealand.
- Lot
- Type/UDI-DI or GTIN/Serial Number: SAM 350P/05060167120671, 05060167127670, +M727SAM350P0/$$+7/17D00023014, 18D00020508, 19D00017655, 20D00006524, 21D91081653, 22D91152391; SAM 360P/05060167127687, 05060167127311,5060167122927/21E91049810, 23E90001379, 23E90010409; SAM 450P/05060167129773, +M727SAM450P0/$$+7, +M727SAM450P0/$$+7/17G00001893, 18G00001553; SAM 500P/05060167122453/22B91071233
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United Kingdom
Related recalls
- FDAHeartSine samaritan PAD (Public Access Defibrillator); HeartSine SAM 350P 350P US Device, US English - 2 MinOctober 7, 2026
- FDAHeartSine SAM 350P 350P US Device, US English - 2 Min CPR UDI-DI codes: 0506016712067 350P US Device, USSeptember 16, 2026
- FDAHeartSine samaritan PAD (Public Access Defibrillator), Type: SAM 350P, SAM 360P, SAM 450PMay 22, 2024
- FDAFluticasone Propionate Nasal Spray USP 50mcg, 120 Metered Sprays - 16 g net fill, Rx Only, Manufactured by:February 28, 2024
- FDAFilmArray Gastrointestinal (GI) Panel RFIT-ASY-0116 (30 test kit) RFIT-ASY-0104 (6 test kit) GastrointestinalFebruary 21, 2024
- FDABIOFIRE Blood Culture Identification 2 (BCID2) Panel when used with: BD BACTEC Lytic Anaerobic medium BDFebruary 21, 2024
Questions about this recall
Was HeartSine samaritan PAD (Public Access Defibrillator): Type/Catalog: SAM 350P/350-UNIT-US-10, 350-STR-US-10, 350-STR-UK-10; SAM 360P/360-STR-JA-10, 360-STR-DE-10, 360-STR-UK-10; SAM 450P/450-UNIT-US-08, ; SAM 500P/500-BAS-UK-10 recalled?
Yes. FDA published recall Z-1141-2024 on February 21, 2024. Reported hazard: Automated external defibrillators were shipped in their test configuration so there is a potential to have incorrect language, and CPR duration resulting in the device prompting users to perform CPR for 5-seconds, and it could analyze heart rhythm and instruct users to deliver a shock every 5-seconds if a shockable heart rhythm is detected, potentially leading to no therapy or delayed therapy.
What should I do if I have this product?
Voluntary: Firm initiated · Letter