Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197
Published February 19, 2020 · Recall initiated: July 19, 2019
Summary
FDA records medical device Class II recall Z-1148-2020, published February 19, 2020 and initiated July 19, 2019. Recalling firm: Stryker Corporation; Status: Terminated; Reported quantity: 87 devices; Lot: 19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.
What to do
Voluntary: Firm initiated · E-Mail
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Stryker Corporation
- Distributor
- US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI
- Lot
- 19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2.
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDASTRYKER PKG., STRYKEFLOW SUCTION/IRRIGATOR PS, Catalog Number 0250070540August 12, 2026
- FDASTRYKER AHTO DUAL SPIKE TUBE SET PACKAGING, Catalog Number 0250070640August 12, 2026
- FDASTRYKER AHTO TUBE SET WITH TIP PACKAGING, Catalog Number 0250070620August 12, 2026
- FDASTRYKER AHTO TUBE SET PACKAGING, Catalog Number 0250070600August 12, 2026
- FDARusch Greenlite Single-Use Metal Laryngoscope Blade for Fiber Optic SystemFebruary 19, 2020
- FDAROSA Brain 3.0 Application-BrainFebruary 19, 2020
Questions about this recall
Was Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197 recalled?
Yes. FDA published recall Z-1148-2020 on February 19, 2020. Reported hazard: Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.
What should I do if I have this product?
Voluntary: Firm initiated · E-Mail