FDAFDA · Food and drugs·Recall number Z-1148-2020

Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197

Published February 19, 2020 · Recall initiated: July 19, 2019

Summary

FDA records medical device Class II recall Z-1148-2020, published February 19, 2020 and initiated July 19, 2019. Recalling firm: Stryker Corporation; Status: Terminated; Reported quantity: 87 devices; Lot: 19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.

What to do

Voluntary: Firm initiated · E-Mail

Get an email when a new FDA recall like this is published

Free. We only write when there's a new recall that matches. Unsubscribe in one click.

Recall details

Manufacturer
Stryker Corporation
Distributor
US: GA, NJ, NY, KY, MD, TX, UT, AZ, IL, ID, CO, OH, FL, AK, CA, NV, AR, ME, WI
Lot
19113AG2, 19130AG2, 19134AG2, 19115AG2, 19131AG2, 19133AG2, 19154AG2,19155AG2, 19156AG2, 19157AG2, 19149AG2.
Risk classification
Class II
Area (FDA)
device
Country
United States

Related recalls

Questions about this recall

Was Stryker, Omega 4.75 mm,. PEEK Knotless Anchor System, Single, REF 3910-500-471, UDI 07613327464818, Sterile R, RX Only CE 0197 recalled?

Yes. FDA published recall Z-1148-2020 on February 19, 2020. Reported hazard: Complaints have been received regarding difficulty in removing the driver from the screw. High driver removal forces can lead to the screw pulling out still attached to the driver. This could result in an additional anchor insertion site in the bone being used.

What should I do if I have this product?

Voluntary: Firm initiated · E-Mail