The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality.
Published February 28, 2024 · Recall initiated: November 15, 2023
Summary
FDA records medical device Class II recall Z-1155-2024, published February 28, 2024 and initiated November 15, 2023. Recalling firm: Canon Medical System, USA, INC.; Status: Ongoing; Reported quantity: 2; Lot: UDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017. Class II means possible temporary or medically reversible effects, or a remote probability of serious health consequences.
Status when published: Ongoing — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.
What's the hazard
VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
What to do
Voluntary: Firm initiated · Letter
Get an email when a new FDA recall like this is published
Free. We only write when there's a new recall that matches. Unsubscribe in one click.
Recall details
- Manufacturer
- Canon Medical System, USA, INC.
- Distributor
- US: CA, LA
- Lot
- UDI-DI: 04987670100208. Serial Number: BGE2292016, BGE2362017
- Risk classification
- Class II
- Area (FDA)
- device
- Country
- United States
Related recalls
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000CNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000FNovember 4, 2026
- FDAALPHENIX INTERVENTIONAL ANGIOGRAPHY SYSTEM REF: INFX-8000VNovember 4, 2026
- FDAAlphenix INFX-8000C/B5October 14, 2026
- FDAThe device is an X-ray system that is capable of radiographic and fluoroscopic studies and is used in anFebruary 28, 2024
- FDAPhilips Allura Xper Interventional Fluoroscopic X-ray Systems, specifically the DIMMs (Dual In-line MemoryFebruary 28, 2024
Questions about this recall
Was The INFX-8000F is designed to take advantage of the latest technological innovations to reduce dose and to save time. A revolutionary graphic user interface and a multi-tasking computer enable the system to fully meet the requirement for optimum image quality, safety, ease of use, advanced efficiency and improved workflow. The advanced floormounted system INFX-8000F uses a new flat panel detector to present fluoroscopic and fluorographic images of high quality. recalled?
Yes. FDA published recall Z-1155-2024 on February 28, 2024. Reported hazard: VAREX changed the manufacturing of the internal board, and the detectors displayed abnormal images. Internal board of the X-ray flat panel detector may display abnormal images, such as horizontal stripes, which may lead to the inability to complete an ongoing exam, or the patient may need to be examined with another x-ray system.
What should I do if I have this product?
Voluntary: Firm initiated · Letter