Back to search
FDAFDA · Food and drugs·Recall number Z-1162-2014

Smith & Nephew RENASYS-AB Abdominal Dressing Kit with Soft Port REF #'s: 66800980 (The RENASYS Soft Port Dressing Kits are used in conjunction with negative pressure wound therapy pumps)

Published March 19, 2014 · Recall initiated: January 9, 2014

Status when published: Terminated — openFDA updates weekly, so this field may not reflect the current status. Always confirm with the official notice.

What's the hazard

Reports of wound fluid and /or blood that were not evacuated from beneath the wound dressing and the pump blockage alarm did not activate has been associated with inadequate Soft Port sealing at the wound site or improper positioning of the Soft Port orifice over the wound contact dressing drape. Inadequate sealing can result in air leaks and contribute to inefficient fluid evacuation from the wo

What to do

Voluntary: Firm initiated

Recall details

Manufacturer
Smith & Nephew Inc.
Distributor
Worldwide Distribution-USA (nationwide) including Puerto Rico and the states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, ID, IL, KY, LA, MA, MS, MT, NC, NJ, NV, NY, OH, OR, PA, TN, TX, VA, WA, WI, WV, HI and the countries Dubai, Curacao, Argentina, Australia, Chile, Germany, France, England, Hong Kong, Kuwait, Lebanon, Libya, Sri Lanka, Malaysia, New Zealand, Manila, Saudi Arabia, South Africa, Singapore, Turkey, Taipei and Vietnam.
Lot
ALL LOT #'s 2013010287 and lower
Risk classification
Class II
Area (FDA)
device
Country
United States